Enter calibration results for work orders to report calibration data for the actual work performed while calibrating a piece of equipment.

Test point data for equipment is copied to the Calibration tab of the Work Orders form for any calibration work order once it is released. See Defining calibration test point data.

If you need to enter calibration results for a work order that you did not specify as being a calibration work order, you must manually enter the test point data.

If a work order is a multiple equipment work order, you can also enter calibration test points and data for child equipment on the work order.

The calibration results you enter determine the Calibration Status of the work order, as well as the Status for each test point.

Generate a calibration record report for a work order as necessary before beginning the calibration process.

To enter calibration results:

  1. Select Work > Work Orders.

  2. Select the work order for which to enter calibration results, and then click the Calibration tab.

    Calibration Status is automatically populated with the status of the calibration and S.O.P., P&ID No., Precision, Increment, P&ID Drawing, and Alert Tolerance based on the information entered on the Calibration tab of the equipment record listed on the work order. Last Calibrated is automatically populated with the completed date for the last calibration work order for the piece of equipment and Calibration Interval with the interval specified on the PM work order.

  3. Select the test point for which to enter calibration results. If the calibration work order is released, Equipment, Sequence, Equipment Org., Test Point, and the unit of measure for the test point are automatically populated based on the test point data entered for the equipment listed on the work order.

  4. Specify this information:

    Equipment - Choose one of the following options if the work order is a multiple equipment work order:

    Option

    Description

    All Equipment

    Specify All Equipment to create the test point for each of the Related Work Orders associated with all of the equipment records on the Calibration tab of the Work Orders form.

    If the work order is a multiple equipment type work order and you click Add Test Point without selecting an existing test point, Equipment is automatically populated with All Equipment. If you select an existing test point and then click Add Test Point on a multiple equipment work order, Equipment, Equipment Description, Equipment Org., Related Work Order, Test Point UOM, Standard UOM, and Output UOM are automatically populated from the selected record.

    If All Equipment is selected for Equipment, then Device Tolerance, Device Reading, and Status are protected, because you cannot distribute these values to all equipment on the work order.

    WO Header Equipment

    Specify WO Header Equipment to create the test point for each equipment record on the Calibration tab of the Work Order on the work order header.

    Equipment

    Specify a specific Equipment record to create the test point for the Related Work Order associated with the selected Equipment.

    See Adding equipment to work orders to split work order costs.

    Not Applicable - Select to indicate the test point is no longer applicable for the calibration process.

    Test Point - Specify the test point value for the calibration, and then enter the unit of measure for the test point in the adjacent field.

    Standard - Specify the standard value for the calibration, and then enter the unit of measure for the standard in the adjacent field.

  5. Specify the following in the As Found section of the form to enter the initial calibration results:

    Device Tolerance - Specify the range of the device tolerance.

    Output - Specify the value of the test point output, and then enter the output’s unit of measure in the adjacent field.

    Device Reading - Specify the reading of the device. Standard is subtracted from the Device Reading to determine the Deviation. Status is populated based on the calibration results you enter.

  6. Specify the following in the As Left section of the form if re-calibration is required:

    Click Copy to As Left to copy the data from the As Found section to the As Left section, and then modify the fields as necessary.

    Specify the calibration Standard, Output, Device Tolerance, and Device Reading. Standard is subtracted from the Device Reading to determine the Deviation. Status is populated based on the calibration results you enter.

  7. Click Submit.

    • To add a new test point, click Add Test Point. Enter data as necessary, and then click Submit.

    • If you delete a test point for which the test point Equipment is associated with a MEC work order, the test point is removed from the Related Work Order.

    • The work order’s calibration status is calculated based on the status of each test point. For multiple equipment work orders, the calibration status of the parent work order considers the test points for all of its related work orders.

    • To run a calibration record report, click Run Calibration Report.